Using low-E resonators to reduce RF heating in biological samples for static solid-state NMR up to 900 MHz

Peter L. Gor'kov, Eduard Y. Chekmenev, Conggang Li, Myriam Cotten, Jarrod J. Buffy, Nathaniel Traaseth, Gianluigi Veglia, William W. Brey

Research output: Contribution to journalArticle

Abstract

RF heating of solid-state biological samples is known to be a destabilizing factor in high-field NMR experiments that shortens the sample lifetime by continuous dehydration during the high-power cross-polarization and decoupling pulses. In this work, we describe specially designed, large volume, low-E 15N-1H solid-state NMR probes developed for 600 and 900 MHz PISEMA studies of dilute membrane proteins oriented in hydrated and dielectrically lossy lipid bilayers. The probes use an orthogonal coil design in which separate resonators pursue their own aims at the respective frequencies, resulting in a simplified and more efficient matching network. Sample heating at the 1H frequency is minimized by a loop-gap resonator which produces a homogeneous magnetic field B1 with low electric field E. Within the loop-gap resonator, a multi-turn solenoid closely matching the shape of the sample serves as an efficient observe coil. We compare power dissipation in a typical lossy bilayer sample in the new low-E probe and in a previously reported 15N-1H probe which uses a double-tuned 4-turn solenoid. RF loss in the sample is measured in each probe by observing changes in the 1H 360° pulse lengths. For the same values of 1H B1 field, sample heating in the new probe was found to be smaller by an order of magnitude. Applications of the low-E design to the PISEMA study of membrane proteins in their native hydrated bilayer environment are demonstrated at 600 and 900 MHz.

Original languageEnglish (US)
Pages (from-to)77-93
Number of pages17
JournalJournal of Magnetic Resonance
Volume185
Issue number1
DOIs
StatePublished - Mar 2007

Fingerprint

Heating
Resonators
resonators
Nuclear magnetic resonance
solid state
nuclear magnetic resonance
heating
Membrane Proteins
probes
Solenoids
Lipid Bilayers
Magnetic Fields
Dehydration
solenoids
coils
Lipid bilayers
membranes
proteins
cross polarization
pulses

Keywords

  • Dielectric loss
  • Field in solenoid
  • Flat-coil probe
  • LGR
  • Lipid bilayers
  • Loop-gap resonator
  • Low-E
  • Membrane protein
  • RF sample heating
  • Solid-state NMR probe

ASJC Scopus subject areas

  • Molecular Biology
  • Physical and Theoretical Chemistry
  • Spectroscopy
  • Radiology Nuclear Medicine and imaging
  • Condensed Matter Physics

Cite this

Using low-E resonators to reduce RF heating in biological samples for static solid-state NMR up to 900 MHz. / Gor'kov, Peter L.; Chekmenev, Eduard Y.; Li, Conggang; Cotten, Myriam; Buffy, Jarrod J.; Traaseth, Nathaniel; Veglia, Gianluigi; Brey, William W.

In: Journal of Magnetic Resonance, Vol. 185, No. 1, 03.2007, p. 77-93.

Research output: Contribution to journalArticle

Gor'kov, Peter L. ; Chekmenev, Eduard Y. ; Li, Conggang ; Cotten, Myriam ; Buffy, Jarrod J. ; Traaseth, Nathaniel ; Veglia, Gianluigi ; Brey, William W. / Using low-E resonators to reduce RF heating in biological samples for static solid-state NMR up to 900 MHz. In: Journal of Magnetic Resonance. 2007 ; Vol. 185, No. 1. pp. 77-93.
@article{f305bc6e57c748a68e530248b7baf076,
title = "Using low-E resonators to reduce RF heating in biological samples for static solid-state NMR up to 900 MHz",
abstract = "RF heating of solid-state biological samples is known to be a destabilizing factor in high-field NMR experiments that shortens the sample lifetime by continuous dehydration during the high-power cross-polarization and decoupling pulses. In this work, we describe specially designed, large volume, low-E 15N-1H solid-state NMR probes developed for 600 and 900 MHz PISEMA studies of dilute membrane proteins oriented in hydrated and dielectrically lossy lipid bilayers. The probes use an orthogonal coil design in which separate resonators pursue their own aims at the respective frequencies, resulting in a simplified and more efficient matching network. Sample heating at the 1H frequency is minimized by a loop-gap resonator which produces a homogeneous magnetic field B1 with low electric field E. Within the loop-gap resonator, a multi-turn solenoid closely matching the shape of the sample serves as an efficient observe coil. We compare power dissipation in a typical lossy bilayer sample in the new low-E probe and in a previously reported 15N-1H probe which uses a double-tuned 4-turn solenoid. RF loss in the sample is measured in each probe by observing changes in the 1H 360° pulse lengths. For the same values of 1H B1 field, sample heating in the new probe was found to be smaller by an order of magnitude. Applications of the low-E design to the PISEMA study of membrane proteins in their native hydrated bilayer environment are demonstrated at 600 and 900 MHz.",
keywords = "Dielectric loss, Field in solenoid, Flat-coil probe, LGR, Lipid bilayers, Loop-gap resonator, Low-E, Membrane protein, RF sample heating, Solid-state NMR probe",
author = "Gor'kov, {Peter L.} and Chekmenev, {Eduard Y.} and Conggang Li and Myriam Cotten and Buffy, {Jarrod J.} and Nathaniel Traaseth and Gianluigi Veglia and Brey, {William W.}",
year = "2007",
month = "3",
doi = "10.1016/j.jmr.2006.11.008",
language = "English (US)",
volume = "185",
pages = "77--93",
journal = "Journal of Magnetic Resonance",
issn = "1090-7807",
publisher = "Academic Press Inc.",
number = "1",

}

TY - JOUR

T1 - Using low-E resonators to reduce RF heating in biological samples for static solid-state NMR up to 900 MHz

AU - Gor'kov, Peter L.

AU - Chekmenev, Eduard Y.

AU - Li, Conggang

AU - Cotten, Myriam

AU - Buffy, Jarrod J.

AU - Traaseth, Nathaniel

AU - Veglia, Gianluigi

AU - Brey, William W.

PY - 2007/3

Y1 - 2007/3

N2 - RF heating of solid-state biological samples is known to be a destabilizing factor in high-field NMR experiments that shortens the sample lifetime by continuous dehydration during the high-power cross-polarization and decoupling pulses. In this work, we describe specially designed, large volume, low-E 15N-1H solid-state NMR probes developed for 600 and 900 MHz PISEMA studies of dilute membrane proteins oriented in hydrated and dielectrically lossy lipid bilayers. The probes use an orthogonal coil design in which separate resonators pursue their own aims at the respective frequencies, resulting in a simplified and more efficient matching network. Sample heating at the 1H frequency is minimized by a loop-gap resonator which produces a homogeneous magnetic field B1 with low electric field E. Within the loop-gap resonator, a multi-turn solenoid closely matching the shape of the sample serves as an efficient observe coil. We compare power dissipation in a typical lossy bilayer sample in the new low-E probe and in a previously reported 15N-1H probe which uses a double-tuned 4-turn solenoid. RF loss in the sample is measured in each probe by observing changes in the 1H 360° pulse lengths. For the same values of 1H B1 field, sample heating in the new probe was found to be smaller by an order of magnitude. Applications of the low-E design to the PISEMA study of membrane proteins in their native hydrated bilayer environment are demonstrated at 600 and 900 MHz.

AB - RF heating of solid-state biological samples is known to be a destabilizing factor in high-field NMR experiments that shortens the sample lifetime by continuous dehydration during the high-power cross-polarization and decoupling pulses. In this work, we describe specially designed, large volume, low-E 15N-1H solid-state NMR probes developed for 600 and 900 MHz PISEMA studies of dilute membrane proteins oriented in hydrated and dielectrically lossy lipid bilayers. The probes use an orthogonal coil design in which separate resonators pursue their own aims at the respective frequencies, resulting in a simplified and more efficient matching network. Sample heating at the 1H frequency is minimized by a loop-gap resonator which produces a homogeneous magnetic field B1 with low electric field E. Within the loop-gap resonator, a multi-turn solenoid closely matching the shape of the sample serves as an efficient observe coil. We compare power dissipation in a typical lossy bilayer sample in the new low-E probe and in a previously reported 15N-1H probe which uses a double-tuned 4-turn solenoid. RF loss in the sample is measured in each probe by observing changes in the 1H 360° pulse lengths. For the same values of 1H B1 field, sample heating in the new probe was found to be smaller by an order of magnitude. Applications of the low-E design to the PISEMA study of membrane proteins in their native hydrated bilayer environment are demonstrated at 600 and 900 MHz.

KW - Dielectric loss

KW - Field in solenoid

KW - Flat-coil probe

KW - LGR

KW - Lipid bilayers

KW - Loop-gap resonator

KW - Low-E

KW - Membrane protein

KW - RF sample heating

KW - Solid-state NMR probe

UR - http://www.scopus.com/inward/record.url?scp=33947605702&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=33947605702&partnerID=8YFLogxK

U2 - 10.1016/j.jmr.2006.11.008

DO - 10.1016/j.jmr.2006.11.008

M3 - Article

VL - 185

SP - 77

EP - 93

JO - Journal of Magnetic Resonance

JF - Journal of Magnetic Resonance

SN - 1090-7807

IS - 1

ER -