Reliability and failure modes of internal conical dental implant connections

Amilcar C. Freitas-Júnior, Erika O. Almeida, Estevam A. Bonfante, Nelson R F A Silva, Paulo Coelho

Research output: Contribution to journalArticle

Abstract

Objective: Biological and mechanical implant-abutment connection complications and failures are still present in clinical practice, frequently compromising oral function. The purpose of this study was to evaluate the reliability and failure modes of anterior single-unit restorations in internal conical interface (ICI) implants using step-stress accelerated life testing (SSALT). Materials and methods: Forty-two ICI implants were distributed in two groups (n = 21 each): group AT-OsseoSpeed TX (Astra Tech, Waltham, MA, USA); group SV-Duocon System Line, Morse Taper (Signo Vinces Ltda., Campo Largo, PR, Brazil). The corresponding abutments were screwed to the implants and standardized maxillary central incisor metal crowns were cemented and subjected to SSALT in water. Use-level probability Weibull curves and reliability for a mission of 50,000 cycles at 200 N were calculated. Differences between groups were assessed by Kruskal-Wallis along with Bonferroni's post-hoc tests. Polarized-light and scanning electron microscopes were used for failure analyses. Results: The Beta (β) value derived from use level probability Weibull calculation was 1.62 (1.01-2.58) for group AT and 2.56 (1.76-3.74) for group SV, indicating that fatigue was an accelerating factor for failure of both groups. The reliability for group AT was 0.95 and for group SV was 0.88. Kruskal-Wallis along with Bonferroni's post-hoc tests showed no significant difference between the groups tested (P > 0.27). In all specimens of both groups, the chief failure mode was abutment fracture at the conical joint region and screw fracture at neck's region. Conclusions: Reliability was not different between investigated ICI connections supporting maxillary incisor crowns. Failure modes were similar.

Original languageEnglish (US)
Pages (from-to)197-202
Number of pages6
JournalClinical Oral Implants Research
Volume24
Issue number2
DOIs
StatePublished - Feb 2013

Fingerprint

Dental Implants
Incisor
Crowns
Psychological Stress
Fatigue
Brazil
Neck
Joints
Metals
Electrons
Light
Water

Keywords

  • Fractography
  • Implant-supported prostheses
  • Internal conical interface connection
  • Step-stress accelerated life testing
  • Weibull

ASJC Scopus subject areas

  • Oral Surgery

Cite this

Reliability and failure modes of internal conical dental implant connections. / Freitas-Júnior, Amilcar C.; Almeida, Erika O.; Bonfante, Estevam A.; Silva, Nelson R F A; Coelho, Paulo.

In: Clinical Oral Implants Research, Vol. 24, No. 2, 02.2013, p. 197-202.

Research output: Contribution to journalArticle

Freitas-Júnior, Amilcar C. ; Almeida, Erika O. ; Bonfante, Estevam A. ; Silva, Nelson R F A ; Coelho, Paulo. / Reliability and failure modes of internal conical dental implant connections. In: Clinical Oral Implants Research. 2013 ; Vol. 24, No. 2. pp. 197-202.
@article{7543557e4bf949ad9c1f5aca448e30a4,
title = "Reliability and failure modes of internal conical dental implant connections",
abstract = "Objective: Biological and mechanical implant-abutment connection complications and failures are still present in clinical practice, frequently compromising oral function. The purpose of this study was to evaluate the reliability and failure modes of anterior single-unit restorations in internal conical interface (ICI) implants using step-stress accelerated life testing (SSALT). Materials and methods: Forty-two ICI implants were distributed in two groups (n = 21 each): group AT-OsseoSpeed™ TX (Astra Tech, Waltham, MA, USA); group SV-Duocon System Line, Morse Taper (Signo Vinces Ltda., Campo Largo, PR, Brazil). The corresponding abutments were screwed to the implants and standardized maxillary central incisor metal crowns were cemented and subjected to SSALT in water. Use-level probability Weibull curves and reliability for a mission of 50,000 cycles at 200 N were calculated. Differences between groups were assessed by Kruskal-Wallis along with Bonferroni's post-hoc tests. Polarized-light and scanning electron microscopes were used for failure analyses. Results: The Beta (β) value derived from use level probability Weibull calculation was 1.62 (1.01-2.58) for group AT and 2.56 (1.76-3.74) for group SV, indicating that fatigue was an accelerating factor for failure of both groups. The reliability for group AT was 0.95 and for group SV was 0.88. Kruskal-Wallis along with Bonferroni's post-hoc tests showed no significant difference between the groups tested (P > 0.27). In all specimens of both groups, the chief failure mode was abutment fracture at the conical joint region and screw fracture at neck's region. Conclusions: Reliability was not different between investigated ICI connections supporting maxillary incisor crowns. Failure modes were similar.",
keywords = "Fractography, Implant-supported prostheses, Internal conical interface connection, Step-stress accelerated life testing, Weibull",
author = "Freitas-J{\'u}nior, {Amilcar C.} and Almeida, {Erika O.} and Bonfante, {Estevam A.} and Silva, {Nelson R F A} and Paulo Coelho",
year = "2013",
month = "2",
doi = "10.1111/j.1600-0501.2012.02443.x",
language = "English (US)",
volume = "24",
pages = "197--202",
journal = "Clinical Oral Implants Research",
issn = "0905-7161",
publisher = "Blackwell Munksgaard",
number = "2",

}

TY - JOUR

T1 - Reliability and failure modes of internal conical dental implant connections

AU - Freitas-Júnior, Amilcar C.

AU - Almeida, Erika O.

AU - Bonfante, Estevam A.

AU - Silva, Nelson R F A

AU - Coelho, Paulo

PY - 2013/2

Y1 - 2013/2

N2 - Objective: Biological and mechanical implant-abutment connection complications and failures are still present in clinical practice, frequently compromising oral function. The purpose of this study was to evaluate the reliability and failure modes of anterior single-unit restorations in internal conical interface (ICI) implants using step-stress accelerated life testing (SSALT). Materials and methods: Forty-two ICI implants were distributed in two groups (n = 21 each): group AT-OsseoSpeed™ TX (Astra Tech, Waltham, MA, USA); group SV-Duocon System Line, Morse Taper (Signo Vinces Ltda., Campo Largo, PR, Brazil). The corresponding abutments were screwed to the implants and standardized maxillary central incisor metal crowns were cemented and subjected to SSALT in water. Use-level probability Weibull curves and reliability for a mission of 50,000 cycles at 200 N were calculated. Differences between groups were assessed by Kruskal-Wallis along with Bonferroni's post-hoc tests. Polarized-light and scanning electron microscopes were used for failure analyses. Results: The Beta (β) value derived from use level probability Weibull calculation was 1.62 (1.01-2.58) for group AT and 2.56 (1.76-3.74) for group SV, indicating that fatigue was an accelerating factor for failure of both groups. The reliability for group AT was 0.95 and for group SV was 0.88. Kruskal-Wallis along with Bonferroni's post-hoc tests showed no significant difference between the groups tested (P > 0.27). In all specimens of both groups, the chief failure mode was abutment fracture at the conical joint region and screw fracture at neck's region. Conclusions: Reliability was not different between investigated ICI connections supporting maxillary incisor crowns. Failure modes were similar.

AB - Objective: Biological and mechanical implant-abutment connection complications and failures are still present in clinical practice, frequently compromising oral function. The purpose of this study was to evaluate the reliability and failure modes of anterior single-unit restorations in internal conical interface (ICI) implants using step-stress accelerated life testing (SSALT). Materials and methods: Forty-two ICI implants were distributed in two groups (n = 21 each): group AT-OsseoSpeed™ TX (Astra Tech, Waltham, MA, USA); group SV-Duocon System Line, Morse Taper (Signo Vinces Ltda., Campo Largo, PR, Brazil). The corresponding abutments were screwed to the implants and standardized maxillary central incisor metal crowns were cemented and subjected to SSALT in water. Use-level probability Weibull curves and reliability for a mission of 50,000 cycles at 200 N were calculated. Differences between groups were assessed by Kruskal-Wallis along with Bonferroni's post-hoc tests. Polarized-light and scanning electron microscopes were used for failure analyses. Results: The Beta (β) value derived from use level probability Weibull calculation was 1.62 (1.01-2.58) for group AT and 2.56 (1.76-3.74) for group SV, indicating that fatigue was an accelerating factor for failure of both groups. The reliability for group AT was 0.95 and for group SV was 0.88. Kruskal-Wallis along with Bonferroni's post-hoc tests showed no significant difference between the groups tested (P > 0.27). In all specimens of both groups, the chief failure mode was abutment fracture at the conical joint region and screw fracture at neck's region. Conclusions: Reliability was not different between investigated ICI connections supporting maxillary incisor crowns. Failure modes were similar.

KW - Fractography

KW - Implant-supported prostheses

KW - Internal conical interface connection

KW - Step-stress accelerated life testing

KW - Weibull

UR - http://www.scopus.com/inward/record.url?scp=84872666532&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84872666532&partnerID=8YFLogxK

U2 - 10.1111/j.1600-0501.2012.02443.x

DO - 10.1111/j.1600-0501.2012.02443.x

M3 - Article

VL - 24

SP - 197

EP - 202

JO - Clinical Oral Implants Research

JF - Clinical Oral Implants Research

SN - 0905-7161

IS - 2

ER -