Abstract

To date, no apparatus has yet been devised which would allow the study of bone microstructure of the whole vertebrae under mechanical loading. This paper outlines the design and development of a 3-axis radio-translucent mechanical testing rig for spinal research and testing. This rig is to be used in conjunction with a Shimadzu micro-CT scanner. Several tests were conducted to verify the feasibility of the rig design. First, the maximum range of deformation in compression, flexion\extension, and lateral bending that could be exerted on a goat lumbar functional spinal unit was evaluated using the noncontact digital markers method. Stepwise compression loading was also conducted on a single porcine vertebra and the loading data was compared to results obtained from an industrial grade compression testing machine. Finally, micro-CT scans of a porcine vertebra prior to and at a compression failure strain were obtained. The rig was confirmed to be able to exert pure moment loading in the above mentioned modes of deformation and the extent of deformation was comparable to previous documented results. The stepwise compression loading conducted in the rig was also found to effectively approximate a continuous loading of the same specimen in an industrial grade compression testing machine. Finally, resultant micro-CT images of isotropic resolution 32.80 μm of a porcine vertebra loaded in the rig were obtained. For the first time, trabecular microarchitecture detail of a whole vertebra buckling under 12.1% failure compression strain loading was studied using voxel-data visualization software. These initial series of tests verify the feasibility of the rig as an apparatus incorporating spinal testing and imaging.

Original languageEnglish (US)
Pages (from-to)957-964
Number of pages8
JournalJournal of Biomechanical Engineering
Volume128
Issue number6
DOIs
StatePublished - Dec 1 2006

Fingerprint

Mechanical testing
Radio
Spine
Compression testing
Swine
Data visualization
Computerized tomography
Testing
Buckling
Bone
Compaction
Goats
Imaging techniques
Software
Microstructure
Bone and Bones
Research

Keywords

  • Bone microarchitecture
  • Micro-CT
  • Radio-translucent
  • Spine
  • Vertebra

ASJC Scopus subject areas

  • Biomedical Engineering
  • Physiology (medical)

Cite this

Radio-translucent 3-axis mechanical testing rig for the spine in micro-CT. / Si-Hoe, K. M.; Teoh, S. H.; Teo, Jeremy.

In: Journal of Biomechanical Engineering, Vol. 128, No. 6, 01.12.2006, p. 957-964.

Research output: Contribution to journalArticle

@article{37c8016ba59d416f9ba4062f47499833,
title = "Radio-translucent 3-axis mechanical testing rig for the spine in micro-CT",
abstract = "To date, no apparatus has yet been devised which would allow the study of bone microstructure of the whole vertebrae under mechanical loading. This paper outlines the design and development of a 3-axis radio-translucent mechanical testing rig for spinal research and testing. This rig is to be used in conjunction with a Shimadzu micro-CT scanner. Several tests were conducted to verify the feasibility of the rig design. First, the maximum range of deformation in compression, flexion\extension, and lateral bending that could be exerted on a goat lumbar functional spinal unit was evaluated using the noncontact digital markers method. Stepwise compression loading was also conducted on a single porcine vertebra and the loading data was compared to results obtained from an industrial grade compression testing machine. Finally, micro-CT scans of a porcine vertebra prior to and at a compression failure strain were obtained. The rig was confirmed to be able to exert pure moment loading in the above mentioned modes of deformation and the extent of deformation was comparable to previous documented results. The stepwise compression loading conducted in the rig was also found to effectively approximate a continuous loading of the same specimen in an industrial grade compression testing machine. Finally, resultant micro-CT images of isotropic resolution 32.80 μm of a porcine vertebra loaded in the rig were obtained. For the first time, trabecular microarchitecture detail of a whole vertebra buckling under 12.1{\%} failure compression strain loading was studied using voxel-data visualization software. These initial series of tests verify the feasibility of the rig as an apparatus incorporating spinal testing and imaging.",
keywords = "Bone microarchitecture, Micro-CT, Radio-translucent, Spine, Vertebra",
author = "Si-Hoe, {K. M.} and Teoh, {S. H.} and Jeremy Teo",
year = "2006",
month = "12",
day = "1",
doi = "10.1115/1.2375136",
language = "English (US)",
volume = "128",
pages = "957--964",
journal = "Journal of Biomechanical Engineering",
issn = "0148-0731",
publisher = "American Society of Mechanical Engineers(ASME)",
number = "6",

}

TY - JOUR

T1 - Radio-translucent 3-axis mechanical testing rig for the spine in micro-CT

AU - Si-Hoe, K. M.

AU - Teoh, S. H.

AU - Teo, Jeremy

PY - 2006/12/1

Y1 - 2006/12/1

N2 - To date, no apparatus has yet been devised which would allow the study of bone microstructure of the whole vertebrae under mechanical loading. This paper outlines the design and development of a 3-axis radio-translucent mechanical testing rig for spinal research and testing. This rig is to be used in conjunction with a Shimadzu micro-CT scanner. Several tests were conducted to verify the feasibility of the rig design. First, the maximum range of deformation in compression, flexion\extension, and lateral bending that could be exerted on a goat lumbar functional spinal unit was evaluated using the noncontact digital markers method. Stepwise compression loading was also conducted on a single porcine vertebra and the loading data was compared to results obtained from an industrial grade compression testing machine. Finally, micro-CT scans of a porcine vertebra prior to and at a compression failure strain were obtained. The rig was confirmed to be able to exert pure moment loading in the above mentioned modes of deformation and the extent of deformation was comparable to previous documented results. The stepwise compression loading conducted in the rig was also found to effectively approximate a continuous loading of the same specimen in an industrial grade compression testing machine. Finally, resultant micro-CT images of isotropic resolution 32.80 μm of a porcine vertebra loaded in the rig were obtained. For the first time, trabecular microarchitecture detail of a whole vertebra buckling under 12.1% failure compression strain loading was studied using voxel-data visualization software. These initial series of tests verify the feasibility of the rig as an apparatus incorporating spinal testing and imaging.

AB - To date, no apparatus has yet been devised which would allow the study of bone microstructure of the whole vertebrae under mechanical loading. This paper outlines the design and development of a 3-axis radio-translucent mechanical testing rig for spinal research and testing. This rig is to be used in conjunction with a Shimadzu micro-CT scanner. Several tests were conducted to verify the feasibility of the rig design. First, the maximum range of deformation in compression, flexion\extension, and lateral bending that could be exerted on a goat lumbar functional spinal unit was evaluated using the noncontact digital markers method. Stepwise compression loading was also conducted on a single porcine vertebra and the loading data was compared to results obtained from an industrial grade compression testing machine. Finally, micro-CT scans of a porcine vertebra prior to and at a compression failure strain were obtained. The rig was confirmed to be able to exert pure moment loading in the above mentioned modes of deformation and the extent of deformation was comparable to previous documented results. The stepwise compression loading conducted in the rig was also found to effectively approximate a continuous loading of the same specimen in an industrial grade compression testing machine. Finally, resultant micro-CT images of isotropic resolution 32.80 μm of a porcine vertebra loaded in the rig were obtained. For the first time, trabecular microarchitecture detail of a whole vertebra buckling under 12.1% failure compression strain loading was studied using voxel-data visualization software. These initial series of tests verify the feasibility of the rig as an apparatus incorporating spinal testing and imaging.

KW - Bone microarchitecture

KW - Micro-CT

KW - Radio-translucent

KW - Spine

KW - Vertebra

UR - http://www.scopus.com/inward/record.url?scp=33846436782&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=33846436782&partnerID=8YFLogxK

U2 - 10.1115/1.2375136

DO - 10.1115/1.2375136

M3 - Article

C2 - 17154698

AN - SCOPUS:33846436782

VL - 128

SP - 957

EP - 964

JO - Journal of Biomechanical Engineering

JF - Journal of Biomechanical Engineering

SN - 0148-0731

IS - 6

ER -