Molecular mass dependence of hyaluronan detection by sandwich ELISA-like assay and membrane blotting using biotinylated hyaluronan binding protein

Han Yuan, Mihir Tank, Abeer Alsofyani, Naman Shah, Nishant Talati, Jaclyn C. LoBello, Jin Kim, Yoji Oonuki, Carol A. De La Motte, Mary Cowman

Research output: Contribution to journalArticle

Abstract

Hyaluronan (HA) is widely detected in biological samples and its concentration is most commonly determined by the use of a labeled specific HA binding protein (aggrecan G1-IGD-G2, HABP), employing membrane blotting and sandwich enzyme-linked immunosorbent assay (ELISA)-like methods. However, the detected signal intensity or the quantified value obtained by using these surface-based methods is related to the molecular mass (M) of HA, especially for HA in the low M range below ~150 kDa. At the same mass or mass concentration, higher M HA gives a higher signal than lower M HA. We have experimentally determined the quantitative relationship between the M of HA (in the range 20-150 kDa) and the relative signal intensity in comparison with a standard HA, in a sandwich ELISA-like assay. An M-dependent signal correction factor (SCF) was calculated and used to correct the signal intensity, so that the corrected concentration value would more accurately reflect the true HA concentration in solution. The SCF for polydisperse low M HA was also calculated and compared with experimental results. When the molecular mass distribution of an HA sample is determined by a method such as gel electrophoresis, then its appropriately averaged SCF can be calculated and used to correct the signal in sandwich ELISA to obtain a more accurate concentration estimation. The correction method works for HA with M between ~150 and 20 kDa, but lower M HA is too poorly detected for useful analysis. The physical basis of the M-dependent detection is proposed to be the increase in detector-accessible fraction of each surface-bound molecule as M increases.

Original languageEnglish (US)
Pages (from-to)1270-1280
Number of pages11
JournalGlycobiology
Volume23
Issue number11
DOIs
StatePublished - Nov 2013

Fingerprint

CD44 Antigens
Immunosorbents
Molecular mass
Hyaluronic Acid
Assays
Enzyme-Linked Immunosorbent Assay
Membranes
Enzymes
Aggrecans
Electrophoresis

Keywords

  • Blotting
  • Detection
  • ELISA
  • Hyaluronan
  • Quantification

ASJC Scopus subject areas

  • Biochemistry

Cite this

Molecular mass dependence of hyaluronan detection by sandwich ELISA-like assay and membrane blotting using biotinylated hyaluronan binding protein. / Yuan, Han; Tank, Mihir; Alsofyani, Abeer; Shah, Naman; Talati, Nishant; LoBello, Jaclyn C.; Kim, Jin; Oonuki, Yoji; De La Motte, Carol A.; Cowman, Mary.

In: Glycobiology, Vol. 23, No. 11, 11.2013, p. 1270-1280.

Research output: Contribution to journalArticle

Yuan, Han ; Tank, Mihir ; Alsofyani, Abeer ; Shah, Naman ; Talati, Nishant ; LoBello, Jaclyn C. ; Kim, Jin ; Oonuki, Yoji ; De La Motte, Carol A. ; Cowman, Mary. / Molecular mass dependence of hyaluronan detection by sandwich ELISA-like assay and membrane blotting using biotinylated hyaluronan binding protein. In: Glycobiology. 2013 ; Vol. 23, No. 11. pp. 1270-1280.
@article{8295b4bfa5614d1bad59b365f0492aca,
title = "Molecular mass dependence of hyaluronan detection by sandwich ELISA-like assay and membrane blotting using biotinylated hyaluronan binding protein",
abstract = "Hyaluronan (HA) is widely detected in biological samples and its concentration is most commonly determined by the use of a labeled specific HA binding protein (aggrecan G1-IGD-G2, HABP), employing membrane blotting and sandwich enzyme-linked immunosorbent assay (ELISA)-like methods. However, the detected signal intensity or the quantified value obtained by using these surface-based methods is related to the molecular mass (M) of HA, especially for HA in the low M range below ~150 kDa. At the same mass or mass concentration, higher M HA gives a higher signal than lower M HA. We have experimentally determined the quantitative relationship between the M of HA (in the range 20-150 kDa) and the relative signal intensity in comparison with a standard HA, in a sandwich ELISA-like assay. An M-dependent signal correction factor (SCF) was calculated and used to correct the signal intensity, so that the corrected concentration value would more accurately reflect the true HA concentration in solution. The SCF for polydisperse low M HA was also calculated and compared with experimental results. When the molecular mass distribution of an HA sample is determined by a method such as gel electrophoresis, then its appropriately averaged SCF can be calculated and used to correct the signal in sandwich ELISA to obtain a more accurate concentration estimation. The correction method works for HA with M between ~150 and 20 kDa, but lower M HA is too poorly detected for useful analysis. The physical basis of the M-dependent detection is proposed to be the increase in detector-accessible fraction of each surface-bound molecule as M increases.",
keywords = "Blotting, Detection, ELISA, Hyaluronan, Quantification",
author = "Han Yuan and Mihir Tank and Abeer Alsofyani and Naman Shah and Nishant Talati and LoBello, {Jaclyn C.} and Jin Kim and Yoji Oonuki and {De La Motte}, {Carol A.} and Mary Cowman",
year = "2013",
month = "11",
doi = "10.1093/glycob/cwt064",
language = "English (US)",
volume = "23",
pages = "1270--1280",
journal = "Glycobiology",
issn = "0959-6658",
publisher = "Oxford University Press",
number = "11",

}

TY - JOUR

T1 - Molecular mass dependence of hyaluronan detection by sandwich ELISA-like assay and membrane blotting using biotinylated hyaluronan binding protein

AU - Yuan, Han

AU - Tank, Mihir

AU - Alsofyani, Abeer

AU - Shah, Naman

AU - Talati, Nishant

AU - LoBello, Jaclyn C.

AU - Kim, Jin

AU - Oonuki, Yoji

AU - De La Motte, Carol A.

AU - Cowman, Mary

PY - 2013/11

Y1 - 2013/11

N2 - Hyaluronan (HA) is widely detected in biological samples and its concentration is most commonly determined by the use of a labeled specific HA binding protein (aggrecan G1-IGD-G2, HABP), employing membrane blotting and sandwich enzyme-linked immunosorbent assay (ELISA)-like methods. However, the detected signal intensity or the quantified value obtained by using these surface-based methods is related to the molecular mass (M) of HA, especially for HA in the low M range below ~150 kDa. At the same mass or mass concentration, higher M HA gives a higher signal than lower M HA. We have experimentally determined the quantitative relationship between the M of HA (in the range 20-150 kDa) and the relative signal intensity in comparison with a standard HA, in a sandwich ELISA-like assay. An M-dependent signal correction factor (SCF) was calculated and used to correct the signal intensity, so that the corrected concentration value would more accurately reflect the true HA concentration in solution. The SCF for polydisperse low M HA was also calculated and compared with experimental results. When the molecular mass distribution of an HA sample is determined by a method such as gel electrophoresis, then its appropriately averaged SCF can be calculated and used to correct the signal in sandwich ELISA to obtain a more accurate concentration estimation. The correction method works for HA with M between ~150 and 20 kDa, but lower M HA is too poorly detected for useful analysis. The physical basis of the M-dependent detection is proposed to be the increase in detector-accessible fraction of each surface-bound molecule as M increases.

AB - Hyaluronan (HA) is widely detected in biological samples and its concentration is most commonly determined by the use of a labeled specific HA binding protein (aggrecan G1-IGD-G2, HABP), employing membrane blotting and sandwich enzyme-linked immunosorbent assay (ELISA)-like methods. However, the detected signal intensity or the quantified value obtained by using these surface-based methods is related to the molecular mass (M) of HA, especially for HA in the low M range below ~150 kDa. At the same mass or mass concentration, higher M HA gives a higher signal than lower M HA. We have experimentally determined the quantitative relationship between the M of HA (in the range 20-150 kDa) and the relative signal intensity in comparison with a standard HA, in a sandwich ELISA-like assay. An M-dependent signal correction factor (SCF) was calculated and used to correct the signal intensity, so that the corrected concentration value would more accurately reflect the true HA concentration in solution. The SCF for polydisperse low M HA was also calculated and compared with experimental results. When the molecular mass distribution of an HA sample is determined by a method such as gel electrophoresis, then its appropriately averaged SCF can be calculated and used to correct the signal in sandwich ELISA to obtain a more accurate concentration estimation. The correction method works for HA with M between ~150 and 20 kDa, but lower M HA is too poorly detected for useful analysis. The physical basis of the M-dependent detection is proposed to be the increase in detector-accessible fraction of each surface-bound molecule as M increases.

KW - Blotting

KW - Detection

KW - ELISA

KW - Hyaluronan

KW - Quantification

UR - http://www.scopus.com/inward/record.url?scp=84885999946&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84885999946&partnerID=8YFLogxK

U2 - 10.1093/glycob/cwt064

DO - 10.1093/glycob/cwt064

M3 - Article

C2 - 23964097

AN - SCOPUS:84885999946

VL - 23

SP - 1270

EP - 1280

JO - Glycobiology

JF - Glycobiology

SN - 0959-6658

IS - 11

ER -