Mechanical testing of implant-supported anterior crowns with different implant/abutment connections.

Erika O. Almeida, Amilcar C. Freitas, Estevam A. Bonfante, Leonard Marotta, Nelson R F A Silva, Paulo G. Coelho

Research output: Contribution to journalArticle

Abstract

This study evaluated the reliability and failure modes of anterior implants with internal-hexagon (IH), external-hexagon (EH), or Morse taper (MT) implant-abutment interface designs. The postulated hypothesis was that the different implant-abutment connections would result in different reliability and failure modes when subjected to step-stress accelerated life testing (SSALT) in water. Sixty-three dental implants (4 × 10 mm) were divided into three groups (n = 21 each) according to connection type: EH, IH, or MT. Commercially pure titanium abutments were screwed to the implants, and standardized maxillary central incisor metallic crowns were cemented and subjected to SSALT in water. The probability of failure versus number of cycles (95% two-sided confidence intervals) was calculated and plotted using a power-law relationship for damage accumulation. Reliability for a mission of 50,000 cycles at 150 N (90% two-sided confidence intervals) was calculated. Polarized-light and scanning electron microscopes were used for failure analyses. The beta values (confidence intervals) derived from use-level probability Weibull calculation were 3.34 (2.22 to 5.00), 1.72 (1.14 to 2.58), and 1.05 (0.60 to 1.83) for groups EH, IH, and MT, respectively, indicating that fatigue was an accelerating factor for all groups. Reliability was significantly different between groups: 99% for MT, 96% for IH, and 31% for EH. Failure modes differed; EH presented abutment screw fracture, IH showed abutment screw and implant fractures, and MT displayed abutment and abutment screw bending or fracture. The postulated hypothesis that different implant-abutment connections to support anterior single-unit replacements would result in different reliability and failure modes when subjected to SSALT was accepted.

Original languageEnglish (US)
Pages (from-to)103-108
Number of pages6
JournalThe International journal of oral & maxillofacial implants
Volume28
Issue number1
StatePublished - Jan 2013

Fingerprint

Crowns
Psychological Stress
Confidence Intervals
Dental Implants
Water
Incisor
Titanium
Fatigue
Electrons
Light

ASJC Scopus subject areas

  • Medicine(all)

Cite this

Mechanical testing of implant-supported anterior crowns with different implant/abutment connections. / Almeida, Erika O.; Freitas, Amilcar C.; Bonfante, Estevam A.; Marotta, Leonard; Silva, Nelson R F A; Coelho, Paulo G.

In: The International journal of oral & maxillofacial implants, Vol. 28, No. 1, 01.2013, p. 103-108.

Research output: Contribution to journalArticle

Almeida, Erika O. ; Freitas, Amilcar C. ; Bonfante, Estevam A. ; Marotta, Leonard ; Silva, Nelson R F A ; Coelho, Paulo G. / Mechanical testing of implant-supported anterior crowns with different implant/abutment connections. In: The International journal of oral & maxillofacial implants. 2013 ; Vol. 28, No. 1. pp. 103-108.
@article{f9e21893146046aaa16c1f400deb8252,
title = "Mechanical testing of implant-supported anterior crowns with different implant/abutment connections.",
abstract = "This study evaluated the reliability and failure modes of anterior implants with internal-hexagon (IH), external-hexagon (EH), or Morse taper (MT) implant-abutment interface designs. The postulated hypothesis was that the different implant-abutment connections would result in different reliability and failure modes when subjected to step-stress accelerated life testing (SSALT) in water. Sixty-three dental implants (4 × 10 mm) were divided into three groups (n = 21 each) according to connection type: EH, IH, or MT. Commercially pure titanium abutments were screwed to the implants, and standardized maxillary central incisor metallic crowns were cemented and subjected to SSALT in water. The probability of failure versus number of cycles (95{\%} two-sided confidence intervals) was calculated and plotted using a power-law relationship for damage accumulation. Reliability for a mission of 50,000 cycles at 150 N (90{\%} two-sided confidence intervals) was calculated. Polarized-light and scanning electron microscopes were used for failure analyses. The beta values (confidence intervals) derived from use-level probability Weibull calculation were 3.34 (2.22 to 5.00), 1.72 (1.14 to 2.58), and 1.05 (0.60 to 1.83) for groups EH, IH, and MT, respectively, indicating that fatigue was an accelerating factor for all groups. Reliability was significantly different between groups: 99{\%} for MT, 96{\%} for IH, and 31{\%} for EH. Failure modes differed; EH presented abutment screw fracture, IH showed abutment screw and implant fractures, and MT displayed abutment and abutment screw bending or fracture. The postulated hypothesis that different implant-abutment connections to support anterior single-unit replacements would result in different reliability and failure modes when subjected to SSALT was accepted.",
author = "Almeida, {Erika O.} and Freitas, {Amilcar C.} and Bonfante, {Estevam A.} and Leonard Marotta and Silva, {Nelson R F A} and Coelho, {Paulo G.}",
year = "2013",
month = "1",
language = "English (US)",
volume = "28",
pages = "103--108",
journal = "International Journal of Oral and Maxillofacial Implants",
issn = "0882-2786",
publisher = "Quintessence Publishing Company",
number = "1",

}

TY - JOUR

T1 - Mechanical testing of implant-supported anterior crowns with different implant/abutment connections.

AU - Almeida, Erika O.

AU - Freitas, Amilcar C.

AU - Bonfante, Estevam A.

AU - Marotta, Leonard

AU - Silva, Nelson R F A

AU - Coelho, Paulo G.

PY - 2013/1

Y1 - 2013/1

N2 - This study evaluated the reliability and failure modes of anterior implants with internal-hexagon (IH), external-hexagon (EH), or Morse taper (MT) implant-abutment interface designs. The postulated hypothesis was that the different implant-abutment connections would result in different reliability and failure modes when subjected to step-stress accelerated life testing (SSALT) in water. Sixty-three dental implants (4 × 10 mm) were divided into three groups (n = 21 each) according to connection type: EH, IH, or MT. Commercially pure titanium abutments were screwed to the implants, and standardized maxillary central incisor metallic crowns were cemented and subjected to SSALT in water. The probability of failure versus number of cycles (95% two-sided confidence intervals) was calculated and plotted using a power-law relationship for damage accumulation. Reliability for a mission of 50,000 cycles at 150 N (90% two-sided confidence intervals) was calculated. Polarized-light and scanning electron microscopes were used for failure analyses. The beta values (confidence intervals) derived from use-level probability Weibull calculation were 3.34 (2.22 to 5.00), 1.72 (1.14 to 2.58), and 1.05 (0.60 to 1.83) for groups EH, IH, and MT, respectively, indicating that fatigue was an accelerating factor for all groups. Reliability was significantly different between groups: 99% for MT, 96% for IH, and 31% for EH. Failure modes differed; EH presented abutment screw fracture, IH showed abutment screw and implant fractures, and MT displayed abutment and abutment screw bending or fracture. The postulated hypothesis that different implant-abutment connections to support anterior single-unit replacements would result in different reliability and failure modes when subjected to SSALT was accepted.

AB - This study evaluated the reliability and failure modes of anterior implants with internal-hexagon (IH), external-hexagon (EH), or Morse taper (MT) implant-abutment interface designs. The postulated hypothesis was that the different implant-abutment connections would result in different reliability and failure modes when subjected to step-stress accelerated life testing (SSALT) in water. Sixty-three dental implants (4 × 10 mm) were divided into three groups (n = 21 each) according to connection type: EH, IH, or MT. Commercially pure titanium abutments were screwed to the implants, and standardized maxillary central incisor metallic crowns were cemented and subjected to SSALT in water. The probability of failure versus number of cycles (95% two-sided confidence intervals) was calculated and plotted using a power-law relationship for damage accumulation. Reliability for a mission of 50,000 cycles at 150 N (90% two-sided confidence intervals) was calculated. Polarized-light and scanning electron microscopes were used for failure analyses. The beta values (confidence intervals) derived from use-level probability Weibull calculation were 3.34 (2.22 to 5.00), 1.72 (1.14 to 2.58), and 1.05 (0.60 to 1.83) for groups EH, IH, and MT, respectively, indicating that fatigue was an accelerating factor for all groups. Reliability was significantly different between groups: 99% for MT, 96% for IH, and 31% for EH. Failure modes differed; EH presented abutment screw fracture, IH showed abutment screw and implant fractures, and MT displayed abutment and abutment screw bending or fracture. The postulated hypothesis that different implant-abutment connections to support anterior single-unit replacements would result in different reliability and failure modes when subjected to SSALT was accepted.

UR - http://www.scopus.com/inward/record.url?scp=84886373412&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84886373412&partnerID=8YFLogxK

M3 - Article

VL - 28

SP - 103

EP - 108

JO - International Journal of Oral and Maxillofacial Implants

JF - International Journal of Oral and Maxillofacial Implants

SN - 0882-2786

IS - 1

ER -