Low temperature degradation and reliability of one-piece ceramic oral implants with a porous surface

Clarisse Sanon, Jérôme Chevalier, Thierry Douillard, Ralf J. Kohal, Paulo Coelho, Jenni Hjerppe, Nelson R F A Silva

Research output: Contribution to journalArticle

Abstract

Low temperature degradation of zirconia (3Y-TZP) oral implants and its effect on fatigue reliability is poorly documented. Objective: The aim of this investigation was to follow the aging process occurring at the surface of implants exhibiting a porous coating and to assess its influence on their mechanical (fatigue) properties. Methods: Tetragonal to monoclinic transformation (t-m) was evaluated during accelerated aging tests up to 100 h in autoclave (134°C, 2 bars) by X-ray diffraction (XRD) and focused ion beam (FIB). A series of implants were steam-aged for 20 h before fatigue testing. Such temperature-time conditions would correspond roughly to 40 years in vivo. The aged specimens and a non-aged control group were step-stress fatigued until failure or survival. Results: The evolution of XRD surface monoclinic content was slow, i.e. 16% and 35% for 20 and 100 h respectively. However, FIB revealed a significant transformation, initiated at the interface between the porous layer and the bulk, preferentially growing towards the bulk. FIB is therefore better indicated than XRD to follow aging in such implants. Higher average fatigue strength (aged 1235 N versus non-aged 826 N) and reliability levels were observed for the 20 h aged group. Significance: After aging for durations compatible with clinical use, 3Y-TZP with porous surface presented higher fatigue performance. This is in contrast to previous studies where loss of strength due to aging was often reported. Generalizations must therefore be avoided when considering aging of zirconia dental products and every new material/process combination should be tested before drawing conclusions.

Original languageEnglish (US)
Pages (from-to)389-397
Number of pages9
JournalDental Materials
Volume29
Issue number4
DOIs
StatePublished - Apr 2013

Fingerprint

Ceramics
Fatigue
Aging of materials
Degradation
Temperature
X-Ray Diffraction
Focused ion beams
Ions
Fatigue of materials
X ray diffraction
Zirconia
Steam
Fatigue testing
Autoclaves
Tooth
Control Groups
Coatings
zirconium oxide

Keywords

  • Cracks
  • Dental implants
  • Fatigue
  • Focused ion beam
  • Reliability
  • Surface modification
  • Transformation
  • Zirconia

ASJC Scopus subject areas

  • Dentistry(all)
  • Materials Science(all)
  • Mechanics of Materials
  • Medicine(all)

Cite this

Sanon, C., Chevalier, J., Douillard, T., Kohal, R. J., Coelho, P., Hjerppe, J., & Silva, N. R. F. A. (2013). Low temperature degradation and reliability of one-piece ceramic oral implants with a porous surface. Dental Materials, 29(4), 389-397. https://doi.org/10.1016/j.dental.2013.01.007

Low temperature degradation and reliability of one-piece ceramic oral implants with a porous surface. / Sanon, Clarisse; Chevalier, Jérôme; Douillard, Thierry; Kohal, Ralf J.; Coelho, Paulo; Hjerppe, Jenni; Silva, Nelson R F A.

In: Dental Materials, Vol. 29, No. 4, 04.2013, p. 389-397.

Research output: Contribution to journalArticle

Sanon, C, Chevalier, J, Douillard, T, Kohal, RJ, Coelho, P, Hjerppe, J & Silva, NRFA 2013, 'Low temperature degradation and reliability of one-piece ceramic oral implants with a porous surface', Dental Materials, vol. 29, no. 4, pp. 389-397. https://doi.org/10.1016/j.dental.2013.01.007
Sanon, Clarisse ; Chevalier, Jérôme ; Douillard, Thierry ; Kohal, Ralf J. ; Coelho, Paulo ; Hjerppe, Jenni ; Silva, Nelson R F A. / Low temperature degradation and reliability of one-piece ceramic oral implants with a porous surface. In: Dental Materials. 2013 ; Vol. 29, No. 4. pp. 389-397.
@article{d7e1124945874a79b786fa15a6a82389,
title = "Low temperature degradation and reliability of one-piece ceramic oral implants with a porous surface",
abstract = "Low temperature degradation of zirconia (3Y-TZP) oral implants and its effect on fatigue reliability is poorly documented. Objective: The aim of this investigation was to follow the aging process occurring at the surface of implants exhibiting a porous coating and to assess its influence on their mechanical (fatigue) properties. Methods: Tetragonal to monoclinic transformation (t-m) was evaluated during accelerated aging tests up to 100 h in autoclave (134°C, 2 bars) by X-ray diffraction (XRD) and focused ion beam (FIB). A series of implants were steam-aged for 20 h before fatigue testing. Such temperature-time conditions would correspond roughly to 40 years in vivo. The aged specimens and a non-aged control group were step-stress fatigued until failure or survival. Results: The evolution of XRD surface monoclinic content was slow, i.e. 16{\%} and 35{\%} for 20 and 100 h respectively. However, FIB revealed a significant transformation, initiated at the interface between the porous layer and the bulk, preferentially growing towards the bulk. FIB is therefore better indicated than XRD to follow aging in such implants. Higher average fatigue strength (aged 1235 N versus non-aged 826 N) and reliability levels were observed for the 20 h aged group. Significance: After aging for durations compatible with clinical use, 3Y-TZP with porous surface presented higher fatigue performance. This is in contrast to previous studies where loss of strength due to aging was often reported. Generalizations must therefore be avoided when considering aging of zirconia dental products and every new material/process combination should be tested before drawing conclusions.",
keywords = "Cracks, Dental implants, Fatigue, Focused ion beam, Reliability, Surface modification, Transformation, Zirconia",
author = "Clarisse Sanon and J{\'e}r{\^o}me Chevalier and Thierry Douillard and Kohal, {Ralf J.} and Paulo Coelho and Jenni Hjerppe and Silva, {Nelson R F A}",
year = "2013",
month = "4",
doi = "10.1016/j.dental.2013.01.007",
language = "English (US)",
volume = "29",
pages = "389--397",
journal = "Dental Materials",
issn = "0109-5641",
publisher = "Elsevier Science",
number = "4",

}

TY - JOUR

T1 - Low temperature degradation and reliability of one-piece ceramic oral implants with a porous surface

AU - Sanon, Clarisse

AU - Chevalier, Jérôme

AU - Douillard, Thierry

AU - Kohal, Ralf J.

AU - Coelho, Paulo

AU - Hjerppe, Jenni

AU - Silva, Nelson R F A

PY - 2013/4

Y1 - 2013/4

N2 - Low temperature degradation of zirconia (3Y-TZP) oral implants and its effect on fatigue reliability is poorly documented. Objective: The aim of this investigation was to follow the aging process occurring at the surface of implants exhibiting a porous coating and to assess its influence on their mechanical (fatigue) properties. Methods: Tetragonal to monoclinic transformation (t-m) was evaluated during accelerated aging tests up to 100 h in autoclave (134°C, 2 bars) by X-ray diffraction (XRD) and focused ion beam (FIB). A series of implants were steam-aged for 20 h before fatigue testing. Such temperature-time conditions would correspond roughly to 40 years in vivo. The aged specimens and a non-aged control group were step-stress fatigued until failure or survival. Results: The evolution of XRD surface monoclinic content was slow, i.e. 16% and 35% for 20 and 100 h respectively. However, FIB revealed a significant transformation, initiated at the interface between the porous layer and the bulk, preferentially growing towards the bulk. FIB is therefore better indicated than XRD to follow aging in such implants. Higher average fatigue strength (aged 1235 N versus non-aged 826 N) and reliability levels were observed for the 20 h aged group. Significance: After aging for durations compatible with clinical use, 3Y-TZP with porous surface presented higher fatigue performance. This is in contrast to previous studies where loss of strength due to aging was often reported. Generalizations must therefore be avoided when considering aging of zirconia dental products and every new material/process combination should be tested before drawing conclusions.

AB - Low temperature degradation of zirconia (3Y-TZP) oral implants and its effect on fatigue reliability is poorly documented. Objective: The aim of this investigation was to follow the aging process occurring at the surface of implants exhibiting a porous coating and to assess its influence on their mechanical (fatigue) properties. Methods: Tetragonal to monoclinic transformation (t-m) was evaluated during accelerated aging tests up to 100 h in autoclave (134°C, 2 bars) by X-ray diffraction (XRD) and focused ion beam (FIB). A series of implants were steam-aged for 20 h before fatigue testing. Such temperature-time conditions would correspond roughly to 40 years in vivo. The aged specimens and a non-aged control group were step-stress fatigued until failure or survival. Results: The evolution of XRD surface monoclinic content was slow, i.e. 16% and 35% for 20 and 100 h respectively. However, FIB revealed a significant transformation, initiated at the interface between the porous layer and the bulk, preferentially growing towards the bulk. FIB is therefore better indicated than XRD to follow aging in such implants. Higher average fatigue strength (aged 1235 N versus non-aged 826 N) and reliability levels were observed for the 20 h aged group. Significance: After aging for durations compatible with clinical use, 3Y-TZP with porous surface presented higher fatigue performance. This is in contrast to previous studies where loss of strength due to aging was often reported. Generalizations must therefore be avoided when considering aging of zirconia dental products and every new material/process combination should be tested before drawing conclusions.

KW - Cracks

KW - Dental implants

KW - Fatigue

KW - Focused ion beam

KW - Reliability

KW - Surface modification

KW - Transformation

KW - Zirconia

UR - http://www.scopus.com/inward/record.url?scp=84875212608&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84875212608&partnerID=8YFLogxK

U2 - 10.1016/j.dental.2013.01.007

DO - 10.1016/j.dental.2013.01.007

M3 - Article

VL - 29

SP - 389

EP - 397

JO - Dental Materials

JF - Dental Materials

SN - 0109-5641

IS - 4

ER -