Implant-abutment connection designs for anterior crowns

Reliability and failure modes

Lucas S. Machado, Estevam A. Bonfante, Rodolfo B. Anchieta, Satoshi Yamaguchi, Paulo Coelho

Research output: Contribution to journalArticle

Abstract

OBJECTIVES: To investigate the effect of implant-abutment connection types on reliability and failure modes of anterior single-unit crowns. METHODS: Fifty-four implants were divided in 3 groups (n = 18 each): external hexagon (EH), internal hexagon (IH), and Morse taper (MT) connection. Abutments were screwed to the implants, and maxillary central incisor metal crowns were cemented and subjected to step-stress accelerated life testing. RESULTS: The beta values derived from use-level probability Weibull calculations for groups IH (2.52), EH (1.67), and MT (0.88) indicated that fatigue influenced the failure only of IH and EH groups. The reliability for a mission of 100,000 cycles at 175 N was 0.99 (0.98-1.00), 0.84 (0.62-0.94) and 0.97 (0.87-0.99) for the EH, IH, and MT, respectively. The characteristic strength was not significantly different between EH (290 N) and IH (251 N) but significantly higher for MT (357 N). For IH and EH groups, failure involved screw fracture, and the MT implants primary failure mode was abutment fracture. CONCLUSIONS: Reliability was higher for the EH and MT relative to IH groups, whereas the characteristic strength was significantly higher for implants with MT connection.

Original languageEnglish (US)
Pages (from-to)540-545
Number of pages6
JournalImplant Dentistry
Volume22
Issue number5
DOIs
StatePublished - Oct 2013

Fingerprint

Crowns
Incisor
Psychological Stress
Fatigue
Metals

Keywords

  • dental implants
  • fractography
  • reliability
  • Weibull

ASJC Scopus subject areas

  • Oral Surgery

Cite this

Implant-abutment connection designs for anterior crowns : Reliability and failure modes. / Machado, Lucas S.; Bonfante, Estevam A.; Anchieta, Rodolfo B.; Yamaguchi, Satoshi; Coelho, Paulo.

In: Implant Dentistry, Vol. 22, No. 5, 10.2013, p. 540-545.

Research output: Contribution to journalArticle

Machado, Lucas S. ; Bonfante, Estevam A. ; Anchieta, Rodolfo B. ; Yamaguchi, Satoshi ; Coelho, Paulo. / Implant-abutment connection designs for anterior crowns : Reliability and failure modes. In: Implant Dentistry. 2013 ; Vol. 22, No. 5. pp. 540-545.
@article{5dd2e008d11947ffa6731c969bc41112,
title = "Implant-abutment connection designs for anterior crowns: Reliability and failure modes",
abstract = "OBJECTIVES: To investigate the effect of implant-abutment connection types on reliability and failure modes of anterior single-unit crowns. METHODS: Fifty-four implants were divided in 3 groups (n = 18 each): external hexagon (EH), internal hexagon (IH), and Morse taper (MT) connection. Abutments were screwed to the implants, and maxillary central incisor metal crowns were cemented and subjected to step-stress accelerated life testing. RESULTS: The beta values derived from use-level probability Weibull calculations for groups IH (2.52), EH (1.67), and MT (0.88) indicated that fatigue influenced the failure only of IH and EH groups. The reliability for a mission of 100,000 cycles at 175 N was 0.99 (0.98-1.00), 0.84 (0.62-0.94) and 0.97 (0.87-0.99) for the EH, IH, and MT, respectively. The characteristic strength was not significantly different between EH (290 N) and IH (251 N) but significantly higher for MT (357 N). For IH and EH groups, failure involved screw fracture, and the MT implants primary failure mode was abutment fracture. CONCLUSIONS: Reliability was higher for the EH and MT relative to IH groups, whereas the characteristic strength was significantly higher for implants with MT connection.",
keywords = "dental implants, fractography, reliability, Weibull",
author = "Machado, {Lucas S.} and Bonfante, {Estevam A.} and Anchieta, {Rodolfo B.} and Satoshi Yamaguchi and Paulo Coelho",
year = "2013",
month = "10",
doi = "10.1097/ID.0b013e31829f1f2d",
language = "English (US)",
volume = "22",
pages = "540--545",
journal = "Implant Dentistry",
issn = "1056-6163",
publisher = "Lippincott Williams and Wilkins",
number = "5",

}

TY - JOUR

T1 - Implant-abutment connection designs for anterior crowns

T2 - Reliability and failure modes

AU - Machado, Lucas S.

AU - Bonfante, Estevam A.

AU - Anchieta, Rodolfo B.

AU - Yamaguchi, Satoshi

AU - Coelho, Paulo

PY - 2013/10

Y1 - 2013/10

N2 - OBJECTIVES: To investigate the effect of implant-abutment connection types on reliability and failure modes of anterior single-unit crowns. METHODS: Fifty-four implants were divided in 3 groups (n = 18 each): external hexagon (EH), internal hexagon (IH), and Morse taper (MT) connection. Abutments were screwed to the implants, and maxillary central incisor metal crowns were cemented and subjected to step-stress accelerated life testing. RESULTS: The beta values derived from use-level probability Weibull calculations for groups IH (2.52), EH (1.67), and MT (0.88) indicated that fatigue influenced the failure only of IH and EH groups. The reliability for a mission of 100,000 cycles at 175 N was 0.99 (0.98-1.00), 0.84 (0.62-0.94) and 0.97 (0.87-0.99) for the EH, IH, and MT, respectively. The characteristic strength was not significantly different between EH (290 N) and IH (251 N) but significantly higher for MT (357 N). For IH and EH groups, failure involved screw fracture, and the MT implants primary failure mode was abutment fracture. CONCLUSIONS: Reliability was higher for the EH and MT relative to IH groups, whereas the characteristic strength was significantly higher for implants with MT connection.

AB - OBJECTIVES: To investigate the effect of implant-abutment connection types on reliability and failure modes of anterior single-unit crowns. METHODS: Fifty-four implants were divided in 3 groups (n = 18 each): external hexagon (EH), internal hexagon (IH), and Morse taper (MT) connection. Abutments were screwed to the implants, and maxillary central incisor metal crowns were cemented and subjected to step-stress accelerated life testing. RESULTS: The beta values derived from use-level probability Weibull calculations for groups IH (2.52), EH (1.67), and MT (0.88) indicated that fatigue influenced the failure only of IH and EH groups. The reliability for a mission of 100,000 cycles at 175 N was 0.99 (0.98-1.00), 0.84 (0.62-0.94) and 0.97 (0.87-0.99) for the EH, IH, and MT, respectively. The characteristic strength was not significantly different between EH (290 N) and IH (251 N) but significantly higher for MT (357 N). For IH and EH groups, failure involved screw fracture, and the MT implants primary failure mode was abutment fracture. CONCLUSIONS: Reliability was higher for the EH and MT relative to IH groups, whereas the characteristic strength was significantly higher for implants with MT connection.

KW - dental implants

KW - fractography

KW - reliability

KW - Weibull

UR - http://www.scopus.com/inward/record.url?scp=84885188046&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84885188046&partnerID=8YFLogxK

U2 - 10.1097/ID.0b013e31829f1f2d

DO - 10.1097/ID.0b013e31829f1f2d

M3 - Article

VL - 22

SP - 540

EP - 545

JO - Implant Dentistry

JF - Implant Dentistry

SN - 1056-6163

IS - 5

ER -